The Laboratory Muffle Furnace YR05284-1 is a state-of-the-art instrument engineered for precision and safety in high-temperature research environments. With a spacious 64-liter chamber measuring 400x400x400 mm, coupled with a sophisticated double-layer structure featuring cooling ventilation, this furnace is designed for safe and efficient operation. Its innovative high-purity alumina fiber insulation ensures optimal energy savings—up to 80% compared to conventional models—while its advanced 30-segment PID controller guarantees precise temperature regulation, safeguarding against overheating and thermocouple failures.
Market Price
The YR05284-1 Muffle Furnace is positioned in the mid-to-high price range within the market, with prices typically ranging from $12,500 to $13,100 USD, depending on added accessories and the country of purchase. This pricing reflects the product’s unbeatable durability, accuracy, and safety features that are highly sought after in research laboratories and industrial applications. For a customized quote and to explore various financing options, we invite you to visit the Kalstein Plus platform, where you can easily generate your quote in just a few steps.
Frequently Asked Questions
- What is the maximum working temperature?
The furnace can handle up to 1700°C, with a recommended continuous working temperature of 1600°C to prolong its lifespan. - What protection features does it offer against overheating?
It includes built-in automatic protection against overheating and continuous thermocouple monitoring to ensure safe operation in controlled environments. - How is temperature control managed?
The furnace utilizes a PID control system featuring 30 programmable segments, ensuring exceptional stability and precision with an accuracy of +/- 1°C.
Advantages and Disadvantages
Key advantages of the YR05284-1 include its durable construction, notable energy savings, and exceptionally precise temperature control. The large-capacity chamber is ideal for handling significant volumes, and its MoSi2 heating system provides uniform heat distribution. However, its weight and dimensions might present challenges in smaller laboratory settings, and for temperatures above 1400°C, a heating rate of just 2°C/min is advised to maintain its durability.
Field Applications
This furnace proves indispensable in sectors such as analytical chemistry, metallurgy, and materials research. It’s perfectly suited for processes involving the heat treatment of ceramics, resistance testing, and sintering—all of which demand stable and controllable temperatures. Thanks to its high precision and enhanced safety features, the YR05284-1 is also the furnace of choice for laboratories engaged in advanced materials testing where every degree counts.
Recommendations for Optimal Use
To extend the furnace’s life, it’s advised to maintain a heating rate under 10°C/min when operating below 1400°C, and to observe recommended rates beyond this threshold. Routine maintenance, encompassing chamber cleaning and thermocouple checks, is crucial for preserving accuracy and safety throughout each operational cycle. Explore the benefits of this professional equipment and enhance your processes with a personalized quote. Through Kalstein Plus, you can efficiently create your quote and receive technical guidance to customize the YR05284-1 Muffle Furnace to your unique needs.
Features
- High-purity alumina fiber insulation for energy efficiency
- Double-layer structure with cooling ventilation for safety
- 30-segment PID controller for precise temperature regulation
- MoSi2 heating elements for uniform heat distribution
- 64-liter chamber capacity for larger volumes
- B type thermocouple
- AC 380V 3 phase or 1 phase 50/60Hz voltage requirement
- Max temperature: 1700 °C, continuous working temp: 1600 °C
Technical Specifications
Model | YR05284-1 |
Rated voltage | AC 380V 3 phase or 1 phase 50/60Hz |
Power | 18kw |
Chamber size | 400x400x400mm, 64L |
Max. Temperature | 1700 °C |
Continuous working temperature | 1600 °C |
Max. Heating Rate | 30 °C /min; Suggest to use 10°C/min under 1400°C, 1400°C-1600°C 5°C/min, >1600°C 2°C/min; |
Temperature controller | PID automatic control and auto-tune function. 30programmable segments for precise control. Built in protection for over-heatingand broken thermocouple |
Temperature Accuracy | +/- 1°C |
Heating Element | MoSi2 heater |
Thermocouple | B type |
Standard parts | Reserve heater 2pcs, crucible tongs 1pc, asbestos gloves 1pair , one user manual |