Biosäkerhetsskåp Klass II A2 YR0090C-SS
Personal · produkt · miljö · trippelt skydd med HEPA H14-filtrering

Kalstein YR0090C-SS är ett Klass II Typ A2 biosäkerhetsskåp för mikrobiologiskt och biomedicinskt arbete med BSL-1 till BSL-3-agens. HEPA H14-filter håller tillbaka 99,995 % av partiklar ≥ 0,3 µm; 70 % av luften recirkuleras och 30 % avleds.
Interactive tour · Explore the YR0090C-SS
Every component, engineered for containment
Six precision-engineered systems working together — press play for a guided tour or click any point on the cabinet to inspect it in detail.
Inspekterbara komponenter
- Dual HEPA filtrationAir filtration
Supply and exhaust HEPA filtration retains particles ≥ 0.3 µm for Class II A2 containment.
- Digital control panelControl
Fan, UV lamp and LED lighting are operated from one digital interface.
- UV disinfection lampDecontamination
Germicidal UV supports chamber decontamination between operations.
- LED work lightingLighting
Waterproof LED lamps provide uniform visibility across the work zone.
- Tempered glass sashSafety
The front sash works as a physical barrier between operator and samples.
- AISI 304 work zoneMaterials
Stainless work surface designed for cleaning and decontamination.
Huvudfunktioner
- Klass II Typ A2 certifierad för BSL-1 till BSL-3
- HEPA H14 nedåtluft- och frånluftsfilter — 99,995 % vid 0,3 µm
- Nedåt 0,33 m/s och inlopp 0,55 m/s för stabil laminär inneslutning
- Arbetskammare i AISI 304 med rundade hörn
- Motoriserad frontruta med säkerhetslarm
- Mikroprocessorstyrning med realtids-luftflödesövervakning
Trippelt skydd · ingenjörslager
Personnel protection
Inflow curtain at 0.55 m/s prevents aerosols from escaping the work zone, isolating the operator from BSL-1 to BSL-3 agents.
- Inflow barrier
- Front sash interlock
- Audible alarms
- Inflow velocity
- 0.55 m/s
- Face velocity
- ≥ 100 fpm
- Max biosafety level
- BSL-3
Product protection
HEPA-filtered downflow at 0.33 m/s bathes the sample in ISO 5 laminar air, keeping the culture free from environmental contamination.
- Downflow laminar
- ISO 5 work zone
- HEPA H14 downflow
- Downflow velocity
- 0.33 m/s
- Work zone class
- ISO 5
- Retention @ 0.3 µm
- 99.995%
Environmental protection
30% of the air is exhausted through a second HEPA H14 filter, releasing only clean air back into the laboratory or ducted exhaust.
- HEPA H14 exhaust
- Top duct connection
- Room air quality preserved
- Exhaust ratio
- 30%
- Exhaust filtration
- HEPA H14
- Noise level
- ≤ 62 dB
Luftflödeslägen · vad som händer i varje riktning
- Downflow — HEPA downflow at 0.33 m/s
Vertical laminar air descends from the supply HEPA over the entire work zone, keeping the sample under ISO 5 conditions.
- Inflow — Inflow curtain at 0.55 m/s
Room air enters through the front grille at 0.55 m/s, creating an air barrier that stops aerosols from escaping.
- Exhaust — HEPA exhaust through the top duct
30% of the internal air is HEPA H14 filtered and released through the top duct.
- Full cycle — Simultaneous triple protection
Downflow, inflow and exhaust combine to protect personnel, product and environment at the same time.
Tekniska specifikationer
| Modell | YR0090C-SS |
|---|---|
| Typ | Class II Type A2 |
| Filtrering | HEPA H14 (99.995% @ 0.3 µm) |
| Nedåtriktad hastighet | 0.33 m/s |
| Inloppshastighet | 0.55 m/s |
| Utsugsfördelning | 30% exhaust · 70% recirculation |
| Yttermått | 1350 × 800 × 2250 mm |
| Arbetskammare | 1130 × 590 × 660 mm |
| Ljudnivå | ≤ 62 dB |
| Belysning | ≥ 1000 lux |
| Strömförsörjning | 220 V / 50-60 Hz · 550 W |
Tillämpningar
- Mikrobiologi och cellodling
- Virologi och BSL-3-forskning
- Klinisk diagnostik och IVF
- Farmaceutisk beredning
- Molekylärbiologi och PCR
Verkliga användningsfall · vägledda flöden
Cell culture and tissue work
Sterile cell culture in a Class II Type A2 environment for cell lines, primary cultures and tissue processing.
- Sanitize the work surface with 70% ethanol
- Start the cabinet and wait for stable airflow
- Transfer materials through the front opening
- Work in the front-to-back laminar zone
- Decontaminate before shutdown
Molecular biology and PCR setup
Prepare PCR reactions and molecular assays free from cross-contamination.
- UV decontaminate for 15 minutes
- Load pipettes, tips and reagents
- Prepare master mix in the clean zone
- Aliquot into reaction tubes
- Clean and log the session
BSL-3 pathogen handling
Manipulation of high-risk agents when integrated in a BSL-3 laboratory suite.
- Verify certification and airflow reports
- Don PPE and follow SOPs
- Introduce material via pass-through
- Perform the procedure inside the containment
- Autoclave waste and decontaminate the chamber
Ideal for applications in
Where the YR0090C-SS performs
Explore each field the YR0090C-SS was engineered for — click any industry to reveal the specific benefits, key metrics and how the cabinet supports the workflow.
Microbiology laboratories
Culture handling, colony isolation and pathogen manipulation under sterile conditions.
key benefits
- Sterile culture handling
- Colony isolation & sub-culturing
- Pathogen manipulation up to BSL-2
- Cleanliness
- ISO 5
- Downflow
- 0.45 m/s
- Continuous
- 24/7
Biotechnology & research
Cell culture, molecular biology and genetic engineering with sample-to-operator isolation.
key benefits
- Primary & stable cell lines
- Transfection / transduction
- CRISPR & genetic engineering
- LED lighting
- 2×10 W
- Stainless
- AISI 304
- Filtration
- Dual HEPA
Pharmaceutical industry
Sterile compounding, formulation R&D and QC of aseptic preparations.
key benefits
- USP 797 aseptic compounding
- Chemotherapy admixtures
- Small-batch sterile filling
- Tipo A2
- Class II
- Recirc / exhaust
- 70/30
- Stainless
- AISI 304
Clinical & diagnostic labs
Processing of blood, tissue and clinical specimens with biological containment.
key benefits
- Blood & serum processing
- Tissue & biopsy handling
- Molecular diagnostics prep
- Containment
- BSL-2
- Decontamination
- UV lamp
- @ 0.3 µm
- ≥99.999%
Biohazardous materials
Safe manipulation of BSL-2 agents, low-volatility chemicals and infectious samples.
key benefits
- BSL-2 infectious agents
- Trace volatile chemicals
- Radionuclide trace work
- HEPA exhaust
- 30%
- Tempered glass
- 5 mm
- Plenum
- Sealed
Fast quotation · 15 seconds
Bring the YR0090C-SS to your laboratory
Receive a full quotation with installation and validation, delivered by Kalstein specialists.
Vanliga frågor
Vilken biosäkerhetsnivå?
Certifierad för BSL-1, BSL-2 och BSL-3 vid installation och validering enligt EN 12469 / NSF 49.
Klass II Typ A2?
Ja. 70 % HEPA-recirkuleras, 30 % HEPA-frånluft via övre kanalen.
Filtrering?
HEPA H14 på nedåt- och frånluft, retention ≥ 99,995 % vid 0,3 µm.
Validering på plats?
Ja, enligt EN 12469 och NSF/ANSI 49.